ORIGINAL RESEARCH
Application of SPE Technique Using a Newly
Obtained Sorbent Based on Silica Gel Covered
with Polyaniline to Simultaneous Determination
of Nitrate (III) and Nitrate (V) Anions
in Water Samples
Ireneusz Sowa, Magdalena Wójciak-Kosior, Ryszard Kocjan
More details
Hide details
Department of Analytical Chemistry, Medical University of Lublin,
Chodźki 4a, 20-093 Lublin, Poland
Pol. J. Environ. Stud. 2013;22(3):881-884
KEYWORDS
ABSTRACT
Nitrate (III) and nitrate (V) often occur together in many environmental samples; however, high concentrations
of other ions, especially chloride, make difficult the correct integration of the peaks. The newly
obtained sorbent, based on silica gel modified with polyaniline, was successfully used as a solid phase in SPE
technique to remove the undesirable matrix effect.
Commercially available bottled water samples with various contents of mineral components after purification
on SPE column were determined using an ion chromatograph DX-500 IC (Dionex USA) with a conductometric
detector. As eluent, aqueous solution containing 3.5 mM (0.37 g/L) of Na2CO3 and 1 mM (0.084
g/L) of NaHCO3 was used. Amounts of anions in all investigated samples of water ranged from 1.71 to 22.33
mg/L for nitrate (V) and 0.04-0.10 mg/L for nitrate (III). These values were within Polish standards.